== Agilent CEカートリッジにC4Dヘッドステージを取り付ける ==
 
== Agilent CEカートリッジにC4Dヘッドステージを取り付ける ==
[http://www.edaq.com/ET120 ET120 C4D ヘッドステージ] はAgilent CE カートリッジに簡単に取り付けることができます。Simply thread the capillary through the channel in the side of the C4D headstage and fix in place.
+
[http://www.edaq.com/ET120 ET120 C4D ヘッドステージ] をAgilent CE カートリッジに取り付けるのは簡単です。C4Dヘッドステージ側面のチャネルにキャピラリーを通して固定するだけです。
  
 
You should ensure that the C4D headstage and its flat grey cable are at least 1 cm away from the capillary as it loops inside the cartridge. This is to prevent the high voltage inside the capillary from shorting through the headstage and its cable, which are both electrically grounded.
 
You should ensure that the C4D headstage and its flat grey cable are at least 1 cm away from the capillary as it loops inside the cartridge. This is to prevent the high voltage inside the capillary from shorting through the headstage and its cable, which are both electrically grounded.

Revision as of 15:40, 13 February 2014

eDAQ C4D システムを Agilent 1600 CE に接続する 

このアプリケーションノートではeDAQ C4D システムをAgilent 1600 CE装置に接続する方法を説明します。

はじめに

eDAQ C4D システムはAgilent 1600 CE装置に接続することができます。

C4D ハードウェアユニットにAgilent検出器を接続する

Agilent 検出器でアナログ出力を持っている機種があります。これにC4D ハードウェアを接続すれば、C4D 及びAgilent検出器からのシグナルを eDAQの PowerChrom ソフトウェア でディスプレイすることができます。

ChemStationでC4Dシグナルを記録する 

Agilent社のChemStation ソフトウェアを使ってC4Dシグナルを記録する場合は、 ER815 C4D 検出器 のアナログ出力を使います。Agilentの1600 CE にはアナログ入力が無いので、 C4Dのような他の検出器と接続する場合は Agilent 35900E データ収録インターフェースが必要です。

Agilent CEカートリッジに取り付けたC4Dヘッドステージ、円形で銀のUV 検出器のセルも確認できます。

Agilent CEカートリッジにC4Dヘッドステージを取り付ける

ET120 C4D ヘッドステージ をAgilent CE カートリッジに取り付けるのは簡単です。C4Dヘッドステージ側面のチャネルにキャピラリーを通して固定するだけです。

You should ensure that the C4D headstage and its flat grey cable are at least 1 cm away from the capillary as it loops inside the cartridge. This is to prevent the high voltage inside the capillary from shorting through the headstage and its cable, which are both electrically grounded.

You should also ensure that the C4D hardware unit is connected to an electrical socket which is grounded. Connect the C4D hardware directly to the electrical socket in the wall, not using an electrical power board.

Citations

  • The eDAQ C4D was used with the Agilent CE for the analysis inorganic ions in drinking water. DOI: 10.1080/10826076.2013.825862
    • The C4D detected both anions and cations in a single injection. This was done by injecting the sample at both ends of the capillary and placing the C4D headstage at an appropriate position along the capillary.
    • It was noted that C4D detection is an alternative and more flexible method of detection compared to the standard optical detection. The effective separation length can easily be adjusted by simply sliding the C4D headstage along the capillary, significantly facilitating the method development and the obtained separation efficiency.
    • The estimated limits of detection (0.07 – 2 ppm) and defined limits of quantification (0.3 – 7 ppm) were comparable or better than those described for indirect UV detection, allowing much broader linear ranges (up to 120 ppm) for most of the studied ions.
  • The Agilent CE was coupled with the eDAQ C4D for the simultaneous determination of atenolol and amiloride in pharmaceutical preparations. Good analytical performance with regards to linearity, reproducibility, and accuracy was achieved using the CE-C4D method. DOI: 10.1002/bmc.1390
  • The determination of cis/trans isomers ofoctadec-9-enoic acid and other long chain fatty acids was performed using the Agilent CE with eDAQ C4D. The method was applied to the determination of fatty acids in margarine samples. The proposed method offers distinct advantages over the GC and HPLC methods, especially in terms of simplicity (without derivatization) and sensitivity. DOI: 10.1016/j.chroma.2013.03.014