== Introduction ==
 +
The eDAQ C4D can be connected to the Agilent 7100 CE instrument, providing contactless conductivity detection for the instrument.
 +
Both the C4D headstage and UV detector can be connected to the capillary at the same time, enabling both types of detection to be performed with a single sample injection.
  
 
== Equipment Required ==
 
== Equipment Required ==
Line 16: Line 20:
  
 
== Choice of Software ==
 
== Choice of Software ==
The data from both C4D and UV-Visible detectors can be recorded and displayed in real time. You can use either the eDAQ PowerChrom software or Agilent ChemStation software. Choose the:
+
The data from both C4D and UV-Visible detectors can be recorded and displayed in real time. You can use either the eDAQ PowerChrom software or Agilent ChemStation software. Use the [Http://www.edaq.com/ET120 ET120 C4D Headstage] and select either the:
* [http://www.edaq.com/ER815 ER815 C4D Detector], can be configured by the user to use either [http://www.edaq.com/ES280 PowerChrom] or third-party software like Agilent ChemStation. The analog output from the ER815 can be connected to the input of the Agilent CE, so the ChemStation software records data from both detectors. The ER815R model includes PowerChrom software to record the C4D signal.
+
* [http://www.edaq.com/ER815 ER815 C4D Detector] can be configured by the user to use either [http://www.edaq.com/ES280 PowerChrom] or third-party software like Agilent ChemStation. The analog output from the ER815 can be connected to the input of the Agilent CE, so the ChemStation software records data from both detectors. The ER815R model includes PowerChrom software to record the C4D signal.
 
* [https://www.edaq.com/ER825 ER825R Two Channel C4D Detector], can also be configured by the user to use either [http://www.edaq.com/ES280 PowerChrom] or third-party software like Agilent ChemStation. The ER825R model includes PowerChrom software and can record the signal from both C4D and UV detectors.
 
* [https://www.edaq.com/ER825 ER825R Two Channel C4D Detector], can also be configured by the user to use either [http://www.edaq.com/ES280 PowerChrom] or third-party software like Agilent ChemStation. The ER825R model includes PowerChrom software and can record the signal from both C4D and UV detectors.
 
* [http://www.edaq.com/ER225 ER225 C4D Data System] can also be used. The analog output of the Agilent UV-Visible detector is connected to the analog input of the ER225. PowerChrom software records the data from both detectors. The ER225 has digital outputs.
 
* [http://www.edaq.com/ER225 ER225 C4D Data System] can also be used. The analog output of the Agilent UV-Visible detector is connected to the analog input of the ER225. PowerChrom software records the data from both detectors. The ER225 has digital outputs.
Line 38: Line 42:
 
You must have [http://www.edaq.com/ES280 PowerChrom software] to do this.
 
You must have [http://www.edaq.com/ES280 PowerChrom software] to do this.
  
You must also have Application firmware installed on the C4D hardware unit. The user can switch the firmware themselves, between Serial and Application firmware, using the ER8x5 Converter and Updater Software (download from [https://www.edaq.com/ES790 here] and find instructions in the [https://www.edaq.com/wiki/C4D_Configurator_Software video] and [https://www.edaq.com/edaq-product-manuals manuals]).
+
You must also have the Application firmware installed on the C4D hardware unit. The user can switch the firmware themselves, between Serial and Application firmware, using the ER8x5 Converter and Updater Software (download from [https://www.edaq.com/ES790 here] and find instructions in the [https://www.edaq.com/wiki/C4D_Configurator_Software video] and [https://www.edaq.com/edaq-product-manuals manuals]).
  
 
Set up the hardware and PowerChrom software as detailed in the [https://www.edaq.com/edaq-product-manuals manuals]. Connect the headstage to the capillary. You should connect the EC027 Trigger cable supplied with the ER815 and ER825, from the CE instrument to the I2C connector at the back of the ER8x5 unit, so that the Agilent ChemStation software sends a message to PowerChrom software to trigger start recording.
 
Set up the hardware and PowerChrom software as detailed in the [https://www.edaq.com/edaq-product-manuals manuals]. Connect the headstage to the capillary. You should connect the EC027 Trigger cable supplied with the ER815 and ER825, from the CE instrument to the I2C connector at the back of the ER8x5 unit, so that the Agilent ChemStation software sends a message to PowerChrom software to trigger start recording.
Line 48: Line 52:
 
=== Using Agilent ChemStation Software ===
 
=== Using Agilent ChemStation Software ===
  
You must also have Serial firmware installed on the C4D hardware unit. The user can switch the firmware themselves, between Serial and Application firmware, using the ER8x5 Converter and Updater Software (download from [https://www.edaq.com/ES790 here] and find instructions in the [https://www.edaq.com/wiki/C4D_Configurator_Software video] and [https://www.edaq.com/edaq-product-manuals manuals]).
+
You must have the Serial firmware installed on the C4D hardware unit. The user can switch the firmware themselves, between Serial and Application firmware, using the ER8x5 Converter and Updater Software (download from [https://www.edaq.com/ES790 here] and find instructions in the [https://www.edaq.com/wiki/C4D_Configurator_Software video] and [https://www.edaq.com/edaq-product-manuals manuals]).
  
 
You should refer to the Agilent 7100 Capillary Electrophoresis System User Manual, which can be downloaded from [https://www.agilent.com here]. The Agilent 7100 Capillary Electrophoresis System has a 20-bit AD converter built in. It enables voltage conversions ranging from 0 V to 1 V.
 
You should refer to the Agilent 7100 Capillary Electrophoresis System User Manual, which can be downloaded from [https://www.agilent.com here]. The Agilent 7100 Capillary Electrophoresis System has a 20-bit AD converter built in. It enables voltage conversions ranging from 0 V to 1 V.

Revision as of 23:15, 30 August 2018

eDAQ C4D connected to the Agilent 7100 CE instrument.
The C4D headstage in the Agilent cartridge. The round, silver UV detector cell can also be seen.

This application note describes how to connect the eDAQ C4D system to the Agilent 7100 CE instrument. See this application note for the Agilent 1600 CE instrument.

Introduction

The eDAQ C4D can be connected to the Agilent 7100 CE instrument, providing contactless conductivity detection for the instrument. Both the C4D headstage and UV detector can be connected to the capillary at the same time, enabling both types of detection to be performed with a single sample injection.

Equipment Required

Choice of Software

The data from both C4D and UV-Visible detectors can be recorded and displayed in real time. You can use either the eDAQ PowerChrom software or Agilent ChemStation software. Use the ET120 C4D Headstage and select either the:

  • ER815 C4D Detector can be configured by the user to use either PowerChrom or third-party software like Agilent ChemStation. The analog output from the ER815 can be connected to the input of the Agilent CE, so the ChemStation software records data from both detectors. The ER815R model includes PowerChrom software to record the C4D signal.
  • ER825R Two Channel C4D Detector, can also be configured by the user to use either PowerChrom or third-party software like Agilent ChemStation. The ER825R model includes PowerChrom software and can record the signal from both C4D and UV detectors.
  • ER225 C4D Data System can also be used. The analog output of the Agilent UV-Visible detector is connected to the analog input of the ER225. PowerChrom software records the data from both detectors. The ER225 has digital outputs.

Installing the C4D Headstage in the Agilent Cartridge

It’s easy to mount the C4D headstage inside the Agilent CE cartridge. Simply thread the capillary through the channel in the side of the C4D headstage and fix in place.

Ensure that you position the C4D headstage near the end of the capillary that is grounded, away from the end of the capillary where the high voltage is applied. This is to ensure that the high voltage doesn’t arc from the background electrolyte inside the capillary, through the thin wall of the capillary, to the headstage and on to the C4D hardware, which could damage the equipment.

You should ensure that the C4D headstage and its flat grey cable are at least 1 cm away from the capillary as it loops inside the cartridge. This is to prevent the high voltage inside the capillary from shorting through the headstage and its cable, which are both electrically grounded.

You should also ensure that the C4D hardware unit is connected to an electrical socket which is grounded.

Installing the ER8x5 Hardware Unit

If you are using the ER815 or ER825 units, you have a choice of recording the data in either PowerChrom software or Agilent ChemStation software.

Using PowerChrom Software

You must have PowerChrom software to do this.

You must also have the Application firmware installed on the C4D hardware unit. The user can switch the firmware themselves, between Serial and Application firmware, using the ER8x5 Converter and Updater Software (download from here and find instructions in the video and manuals).

Set up the hardware and PowerChrom software as detailed in the manuals. Connect the headstage to the capillary. You should connect the EC027 Trigger cable supplied with the ER815 and ER825, from the CE instrument to the I2C connector at the back of the ER8x5 unit, so that the Agilent ChemStation software sends a message to PowerChrom software to trigger start recording.

Selecting C4D Settings The C4D settings (amplitude, frequency and headstage gain) should be selected in the PowerChrom software. Do not use the C4D Configurator Software when the ER815 and ER825 are configured with applications firmware, or when using the ER225.

CE-part configuration.

Using Agilent ChemStation Software

You must have the Serial firmware installed on the C4D hardware unit. The user can switch the firmware themselves, between Serial and Application firmware, using the ER8x5 Converter and Updater Software (download from here and find instructions in the video and manuals).

You should refer to the Agilent 7100 Capillary Electrophoresis System User Manual, which can be downloaded from here. The Agilent 7100 Capillary Electrophoresis System has a 20-bit AD converter built in. It enables voltage conversions ranging from 0 V to 1 V.

Signal Cable You should connect the analog output from the C4D hardware (“Delta K” BNC connector at the back of the ER8x5 unit) to the analog input of the Agilent 7100 CE instrument, using a BNC to BNC cable.

Setting Up ChemStation software Refer to the ChemStation manual. You must tell the software to accept a signal from an external detector. This is done from the dialog window shown in the "CE-part configuration" figure, which is accessible in the software under Instrument/configuration, CE-part configuration. In the "Analog In Signal" tick the "Acquire" box. Give it a name, like "C4D". Change “Signal unit” to V. Set the Multiplier to 1000 (this should mean that the software can record up to 1 V). Then in the eDAQ C4D Configurator software, set the Full Scale Output to 1V. This should mean that both the Agilent analog input and the eDAQ analog output are both set to 1 V.

Selecting C4D Settings The C4D Configurator software allows the setting up of the C4D settings (amplitude, frequency and headstage gain) of ER815 and ER825 C4D Detector when they are configured with the serial firmware. It provides a simple alternative to the use of direct serial commands. Download the C4D Configurator software from the Software Downloads. The procedure for using this software is shown in a video and has been written up as a pdf manual.

Citations

  • The eDAQ C4D was used with the Agilent CE for the analysis inorganic ions in drinking water. DOI: 10.1080/10826076.2013.825862
    • The C4D detected both anions and cations in a single injection. This was done by injecting the sample at both ends of the capillary and placing the C4D headstage at an appropriate position along the capillary.
    • It was noted that C4D detection is an alternative and more flexible method of detection compared to the standard optical detection. The effective separation length can easily be adjusted by simply sliding the C4D headstage along the capillary, significantly facilitating the method development and the obtained separation efficiency.
    • The estimated limits of detection (0.07 – 2 ppm) and defined limits of quantification (0.3 – 7 ppm) were comparable or better than those described for indirect UV detection, allowing much broader linear ranges (up to 120 ppm) for most of the studied ions.
  • The Agilent CE was coupled with the eDAQ C4D for the simultaneous determination of atenolol and amiloride in pharmaceutical preparations. Good analytical performance with regards to linearity, reproducibility, and accuracy was achieved using the CE-C4D method. DOI: 10.1002/bmc.1390
  • The determination of cis/trans isomers ofoctadec-9-enoic acid and other long chain fatty acids was performed using the Agilent CE with eDAQ C4D. The method was applied to the determination of fatty acids in margarine samples. The proposed method offers distinct advantages over the GC and HPLC methods, especially in terms of simplicity (without derivatization) and sensitivity. DOI: 10.1016/j.chroma.2013.03.014