== Introduction ==
 
== Introduction ==
 
eDAQ has developed a C4D headstage for use with capillaries with an outer diameter of 150 μm. The headstage was used for the analysis of a solution containing potassium, sodium and lithium chlorides dissolved in MES/His background electrolyte (BGE).
 
eDAQ has developed a C4D headstage for use with capillaries with an outer diameter of 150 μm. The headstage was used for the analysis of a solution containing potassium, sodium and lithium chlorides dissolved in MES/His background electrolyte (BGE).
The headstage can be used with either the C4D Amp or  C4D Detector.
+
The headstage can be used with any eDAQ C4D hardware.
  
 
A limit of detection of 3 µM was obtained.
 
A limit of detection of 3 µM was obtained.
Line 21: Line 21:
  
 
== eDAQ Equipment Required ==
 
== eDAQ Equipment Required ==
* [Http://www.edaq.com/ER225 ER225 C4D Data System], or [Http://www.edaq.com/ER283 ER283 C4D System Bundle], or [Http://www.edaq.com/ER125 ER125 C4D Detector]
+
* C4D hardware unit, either:
 +
** [Http://www.edaq.com/ER225 ER225 C4D Data System], or
 +
** [http://www.edaq.com/ER815 ER815 C4D Detector] with a third-party data acquisition system
 
* [Http://www.edaq.com/ET120 ET120/150 C4D Headstage (150 μm)]
 
* [Http://www.edaq.com/ET120 ET120/150 C4D Headstage (150 μm)]

Latest revision as of 14:58, 23 January 2014

The eDAQ 150 μm C4D Headstage
Three cations

A C4D (capacitively-coupled contactless conductivity detector) headstage is described for use with 150 μm outer diameter capillaries and for the analysis of three cations.

Introduction

eDAQ has developed a C4D headstage for use with capillaries with an outer diameter of 150 μm. The headstage was used for the analysis of a solution containing potassium, sodium and lithium chlorides dissolved in MES/His background electrolyte (BGE). The headstage can be used with any eDAQ C4D hardware.

A limit of detection of 3 µM was obtained.

Experimental

  • Capillary from Polymicro Technologies, 150 µm OD, 76 µm ID, 60 cm long.
  • Sample: 200 µM Na, K, Li (as chloride salts) dissolved in water and BGE
  • BGE: 20 mM of 2-(N-morpholino)ethanesulfonic acid and L-histidine
  • The new capillary was prepared by flushing with 0.1 M HCl, water, 0.1 M NaOH, water, then BGE, for 5 minutes each
  • Instrument: built in-house at University of Basel
  • Injection: 10 s siphoning at 10 cm height
  • Separation voltage: 25 kV (10 μA current)
  • C4D settings: 700 kHz, 100 V, gain ON

eDAQ Equipment Required