Difference between revisions of "Reference Electrode Potentials"

From eDAQ Wiki
Jump to: navigation, search
(The Calomel Electrode)
(The Calomel Electrode)
Line 3: Line 3:
 
Hg<sub>2</sub >Cl<sub>2</sub> + 2e<sup>–</sup> → 2Hg<sub><i>liq</i></sub> + 2Cl<sup>–</sup>
 
Hg<sub>2</sub >Cl<sub>2</sub> + 2e<sup>–</sup> → 2Hg<sub><i>liq</i></sub> + 2Cl<sup>–</sup>
  
As this equation implies, the electrode potential is dependent on chloride concentration, but independent of hydrogen ion (acid) concentration.  
+
As this equation implies, the electrode potential is dependent on chloride concentration, but independent of hydrogen ion (acid) concentration. Commercial calomel electrodes are available from:
 +
 
 +
: [http://www.koslow.com] Koslow Scientific (USA)
 +
: [http://www.ionode.com] Ionode Pty Ltd (Australia)
 +
 
 +
In Europe the use of calomel electrodes is increasingly problematic because many countries no longer permit the use of mercury-containing devices.
 
  
 
  
  

Revision as of 14:04, 4 August 2014

The Calomel Electrode

The calomel electrode is usually constructed from a platinum wire inserted into a mixture of calomel (mercurous chloride, Hg2Cl2) and liquid mercury, with an electrolyte solution of KCl or NaCl. The relevant half cell equation is: Hg2Cl2 + 2e → 2Hgliq + 2Cl

As this equation implies, the electrode potential is dependent on chloride concentration, but independent of hydrogen ion (acid) concentration. Commercial calomel electrodes are available from:

[1] Koslow Scientific (USA)
[2] Ionode Pty Ltd (Australia)

In Europe the use of calomel electrodes is increasingly problematic because many countries no longer permit the use of mercury-containing devices.


Table 1. Potential (V) of the Calomel Electrode at 25°C
Conditions vs NHE vs SCE LJ Reference
Hg/Hg2Cl2, KCl (0.1 M) 0.3337 0.0925 - 1, 3
Hg/Hg2Cl2, KCl (0.1 M) 0.336 0.092 Yes 2
NCE 0.2801 0.0389 - 1, 3
NCE 0.283 0.039 Yes 2
Hg/Hg2Cl2, KCl (3. 5M) 0.250 0.006 Yes 2
SCE 0.2412 0 - 1, 3
SCE 0.244 0 Yes 2
SSCE 0.2360 -0.0052 - 1

Notes

LJ, liquid junction. Value obtained using a cell which included a liquid junction potential.
NCE, normal calomel electrode: Hg/Hg2Cl2, KCl (1 M)
NHE, normal hydrogen electrode
SCE, saturated calomel electrode: Hg/Hg2Cl2, KCl (sat'd)
SSCE, saturated salt calomel electrode: Hg/Hg2Cl2, NaCl (sat'd)
For values at other temperatures see a calculator here.


References

1. "Electrochemical Methods: Fundamentals and Applications", AJ Bard and LR Faulkner, John Wiley & Sons, NY (2000). See the table on inside back cover.
2. "Electrochemistry for Chemists, Second Edition", DT Sawyer, AJ Sobkowiak, J Roberts, Jr., John Wiley & Sons, NY (1995). See Section 5.2.
3. "Handbook of Analytical Chemistry", L Meites (ed.), McGraw Hill, NY (1963). See Section 5.