Signal Sources and Amplifiers

From eDAQ Wiki
Jump to: navigation, search

Introduction

In order to record an experimental signal with high fidelity 3 issues need to be considered.

  • Signal Sources to be measured
  • Characteristics of the signal sources
  • Type of Amplifier used to record the signal

Signal Sources

Generalised Signal source

There are 4 types or classes of electrical signal sources which occur in various experimental situations:

  • Single ended - Floating
  • Single ended - Grounded with Common mode voltage
  • Balanced - Floating
  • Balanced - Grounded with Common mode voltage

It is important to understand the characteristics of these signal sources in order that they can be matched to the correct amplifier/recording system without degrading the accuracy of the measurement.

NOTE: In discussing the characteristics of signal sources reference will be made to "Source Impedance" which in its most general form consists of a combination of 3 electrical elements: Resistance,Inductance and Capacitance. In most cases one of these elements will predominate, most commonly resistance. So when the term "Source Resistance" is used it implies that the impedance of the source is mainly resistive in nature. At DC or low frequencies source impedance is mainly resistive. The other elements, capacitance and inductance only come into play at higher frequencies.

http://en.wikipedia.org/wiki/Electrical_impedance

In its most general form a signal source can be described by the diagram Generalised Signal source


The 4 classes are listed and discussed below:

Single Ended - Floating

SE Floating

Single Ended Floating signals appears between the two terminals of the signal source. Either of these terminals can be grounded without altering the potential across the two terminals. All signal sources can be regarded as Voltage or Current sources as defined in Nortons Theorem.

http://en.wikipedia.org/wiki/Norton%27s_theorem.

In an analytical laboratory:

Typical voltage ranges: 1 uV (thermocouples) to 30,000 volts (capillary electrophoresis high voltage supplies) - internal impedances are typically low.

Typical current ranges: <1 pA (small amperometric electrode) to >10 amps (high power potentiostat) - internal impedances are typically high.

Signal frequency content: DC to MHz

An ideal voltage source has a near zero internal impedance. An accurate reading can only be obtained with a measuring circuit whose own internal resistance is much higher than the source resistance of the source. Voltmeters therefore need to have a high internal resistance - ideally infinite.

An ideal current source has a near infinite internal impedance. In this case to measure the source current the measurement impedance should be very low. Ammeters therefore need to have a low internal resistance - ideally zero.

Examples

SE Voltage sources

  • Battery - very low internal resistance
  • pH cell - very high internal resistance
  • Thermocouple - very low internal resistance

SE Current sources

  • Silicon detector - high internal resistance
  • Electrochemical sensor - high internal resistance
  • Piezo-Electric sensor - very high resistance


Single ended - Grounded with Common mode voltage

In this configuration one of the terminals is effectively connected to ground via another signal source - called a "Common Mode Voltage". This CM Voltage may be intentional but in most cases it is unintentional and can present a significant problem in the measurement of the signal of interest. In this case the effective signal occurs across the two terminals of the source and is made up of the signal with the CM voltage added to it as shown below. A special but common case of this configuration is when the CM voltage is zero and one of the terminals is effectively connected to ground.

Examples of Single Ended Grounded with common mode

  • AC Line powered signal generator
  • Resistive shunt for measuring current
  • AC powered transducer
  • Chromatography detector


Balanced - Floating

In this configuration the two terminals are both active and have equal impedance to a common point which is floating and unconnected to ground. The voltage on the two terminals moves in different directions in relation to the common point.

Examples of Balanced floating Source

  • Four Arm Wheatstone Bridge with floating power supply
  • Linear variable differential transformer (LVDT)
  • Center tapped Transformer


Balanced - Grounded with Common mode voltage

In this configuration the two terminals are both active and have equal impedance to a common point which is connected to ground via a CM voltage. The voltage on the two terminals moves in different directions in relation to the common point. The CM voltage can be zero when the center point is directly grounded, it can be at a fixed DC value or can result from unintentional connection to ground.

Examples of Balanced floating Source with common mode voltage

  • Four Arm Wheatstone Bridge with line powered supply
  • Differential output Operational Amplifier


Source Characteristics

Characteristics of electrical signal sources in a chemistry analytical laboratory.

Examples:

Single Ended Referenced to Ground

In this case one terminal contains the signal and the other terminal is referenced to ground. This reference to ground can be modelled by an Impedance and a series voltage source. These are referred to as Common Mode Impedance (Zcm) and Common Mode voltage (Vcm)

There are three distinct conditions:

  • Zcm is infinite: The signal source becomes a “Floating source”. Examples include pH electrode, pH/ion selective electrode, piezo electric sensor, thermocouple, and battery or battery powered instrument or meter.
  • Zcm is zero: The signal source becomes “Grounded source”. One terminal carries the signal and the other is connected to ground or 0 volts. An example is an AC powered instrument like a signal generator.
  • Zcm is nonzero: This represents most practical situations. This term is sometimes introduced as a necessary component in the measurement system but typically it represents a “noise” element to be avoided by the choice of an appropriate amplifier input configuration.


AC powered equipment

Source Impedance

The signal source impedance will determine the type of Amplifier required to measure the signal.

Source Voltage

The source voltage will determine the amplification level required in the recorder amplifier

Source Frequency content

The source frequency content will determine the frequency response of the recorder amplifier.

Amplifier types

The amplifier input configuration required will be determined by the Source configuration discussed above.

Single ended - Grounded

Single ended - Isolated

Balanced - Grounded

Balanced - Isolated