Zahner ZENNIUM E Electrochemical Workstation

profile_overview_image

Electrochemical Workstation Zennium E

The Zennium E is a high-performance electrochemical workstation at a surprising affordable price. It provides high-end features like frequency range up to 2 MHz, max. current of ±2.0 A, a controlled voltage of ±12 V and an input impedance of 10 TΩ || 5 pF. The Zennium E is a perfect workstation for a wide range of electrochemical applications with the focus on corrosion, coatings and laminates.

It can be extended by a number of ZAHNER probes.

 

Zennium E was developed using our 35 years of experience in producing high-precision electrochemical workstations of the high-end class. The Zennium E is a revolution in compact electrochemical workstations. It provides a wide frequency range from 10 μHz to 2 MHz, an output current up to ±2.0 A and fast signal processing. Special measurement techniques guarantee an ultra high accuracy and a minimal interference with the test object. The Zennium E can be easily switched from grounded to floating operation.

Zennium E comes bundled with the outstanding Thales Z 3 software package which offers all standard methods and more at a mouse click. This is why the Zennium E can easily be adapted to very different measurement requirements. Furthermore, with the probes available, the Zennium E is able to grow with its tasks:

  • Compliance Voltage up to ±120 V (CVB120)
  • Input Impedance up to 1000 TΩ (HiZ Probe)
  • FRA interface to 3rd-party electronic loads or potentiostats (FRA Probe)

Photos [ click on the thumbnails below to view large images ]

Compact Electrochemical Workstation Zennium E

General

Overall bandwidth

DC - 2.5 MHz

ADC resolution

3 ADCs @ 18 bit

Harmonic reject

> 60 dB @ ½ full scale

Potentiostat modes

Potentiostatic, galvanostatic, pseudo-galvanostatic, rest potential, off, ZRA, FRA

Cell connection

2-, 3- ,4-terminal kelvin

Floating / grounded

Switchable

Chassis

Ground

PC interface

USB

Dimensions

255 x 160 x 385 mm

Weight

8 kg

Accessories

Cell cable set, USB cable, power cord, manual

Power supply

230/115/100VAC, 50/60Hz

Ambient temperature, humidity

+15 °C to +35 °C, < 50% without derating

 

Frequency generator & analyzer

Frequency range

10 µHz to 2 MHz

Accuracy

< 0.0025%

Resolution

0.0025%, 10000 steps/decade

 

Output potentiostatic

Controlled voltage

±4 V, ±12 V

 

Resolution

125 µV, 375 µV

 

Accuracy

±500 µV, ±2 mV / ±0.025% of set value

 

Temperature stability

better 20 µV / °C

 

Compliance voltage

±14 V
±120 V

Main
CVB120

AC amplitude

1 mV to 1 V
1 mV to 25 V

Main
CVB120

Bandwidth

2 MHz @ 33 Ω load

 

IR compensation

auto AC impedance technique
range 0 to 10 MΩ
resolution 0.012%



Small signal rise time

250 ns to 200 µs in 5 steps, automatic selection  by automatic stability control

 

Slew rate

15 MV/s

 

Phase shift

10° @ 250 kHz

 

 

Output galvanostatic

Full scale current ranges

±100 nA to ±2.0 A in 10 steps, resolution 0.0031 % (16 bit) of range
lowest full scale range ±100 nA, resolution 12.5 pA

Main

 

±1 nA to ±0.5 A in 12 steps, resolution 0.0031 % (16 bit) of range
lowest full scale range ±1 nA, resolution 125 fA

HiZ

Accuracy **

±0.1 % of set value @ >|2 µA| to |100 mA|
±1 % of set value @ |1 nA| to |2 µA| or >|100 mA|
±1 % of set value, ±20 pA @  <|1 nA|

Main

 

±1 % of set value, ±250 fA @  < |1 nA|

HiZ

 

Input

Full scale potential ranges

±1, ±2, ±4 V, ±12 V

 

Potential resolution DC *

0.0008 % / 32 µV (4 V range)
0.0008 % / 100 µV (12 V range)

 

Potential resolution AC *

16 nV

 

Potential accuracy DC **

±0.025 % of reading ±0.25 mV (4 V range)
±0.025 % of reading ±1 mV (12 V range)

 

Offset temperature stability

< 10 µV / °C

 

Full scale current range *

±100 pA to ±2.0 A in 33 ranges,
automatic range selection
±1 pA to ±0.5 A in 35 ranges,
automatic range selection

Main

HiZ

Current accuracy DC **

±0.05 % of reading @ > |2 µA| to |100 mA|
±0.5 % of reading @ < |2 µA| or > |100 mA|
±0.5 % of reading, ±10 pA @ < |1 nA|

Main

 

±0.5 % of reading, ±125 fA @ < |1 nA|

HiZ

Input bias current **

|±1 pA| (typ.) / |±5 pA| (max.)
|±10 fA| (typ.) / |±50 fA| (max.)

Main
HiZ

Current Resolution DC *

2.5 pA
25 fA

Main
HiZ

Current Resolution AC *

1.6 fA
16 aA

Main
HiZ

Input impedance

10 TΩ || ±5 pF (typ.)
1000 TΩ || ±1 pF (typ.)

Main
HiZ

Impedance range

100 mΩ to 10 MΩ / 0.2%
1 mΩ to 1 GΩ / 2%
30 µΩ to 1 GΩ / 2%
100 mΩ to 100 GΩ / 3%

Main
Pot
Gal
HiZ

Common mode rejection

> 86 dB @ 10 µHz to 100 kHz
> 66 dB @ 100 kHz to 2 MHz

 

Input channel phase-tracking accuracy

±0.1 ° @ 10 µHz to 100 kHz
±0.25 ° @ 100 kHz to 2 MHz

 

Equivalent effective input noise

2 µV rms / 200 fA rms @ 1 mHz to 10 Hz
20 µV rms / 30 fA rms @ 1 mHz to 10 Hz

Main
HiZ

*  Garanteed by design. For details refer to application note "how to read specifications".

** In the first 6 months after factory calibration, after 20 min. of warm-up.