EPU452  Quad Multi Function isoPod™ with USB



EPU452 Quad Multi Function isoPod™  with USB
EPU452 Quad Multi Function isoPod™ with USB
 
  • Up to four channels available
  • Each channel can be configured for a different sensor
  • Electrical isolation on each channel
  • USB plug and play connectivity
  • Customise operation with your own LabVIEW applications or use Pod-Vu software (included)

Available in a Special Kit including 5 electrodes and software  

 
Unit Price: $ (USD) 3265.00

 

Compatible with Multiple Probes

Using its MF Configuator software, each channel can be independently configured and calibrated for a particular type of sensor:

  • pH, ion selective electrode (ISE), Oxidation Reduction Potential (ORP)
  • Temperature (1000 ohm RTD or 30 kohm thermistor)
  • Dissolved oxygen (dO2)
  • Solution conductivity
  • Amperometric biosensors
  • Voltage (mV)

Thus the instrument can have four pH channels, or one each of pH, conductivity, dO2 and temperature, or any other combination!

Pod-Vu Software Included

Online monitoring of the signals from the different channels, using Windows XP or later computers, is facilitated by Pod-Vu software. You can record signals from one or two Quad MF isoPod units (that is up to eight signals simultaneously).

Customise with your Own Software

You can also write your own recording or process control software using  the 'virtual serial protocol' provided over the isoPod's USB interface.  The serial protocol can be accessed by:

  • Writing your own software, for example in LabVIEW, C#, Visual Basic, etc.
  • Terminal emulation software compatible with serial (COM) ports, eg TeraTermHyperTerminal.
  • Other serial (COM) port compatible software that can log responses into a file or Excel document, eg WinWedgeHyperAccess.

OS Compatibility

The EPU452 is a multi-platform device that uses a 'USB CDC serial port' which is defined in the USB standard. Thus it is compatible with any operating system that supports a USB CDC serial port. This includes Windows, Mac OS X, and Linux. Please note, however, that MF Configurator and Pod-Vu software are only available for Windows XP and later computers.

Photos[ click on the thumbnails below to view large images ]

EPU452 with conductivity, RTD temperature, dissolved oxygen and pH electrodes EPU452 with conductivity, RTD temperature, dissolved oxygen and pH electrodes EPU452 with conductivity, RTD temperature, dissolved oxygen and pH electrodes

Videos

The isoPods for Recording Temperature, pH, Conductivity and Dissolved Oxygen
 
Serial Port Monitor for hardware which use serial ports (isoPods, HVS)
 

View more related videos on the Screencast Training Videos Wiki

 
General
  • Input connectors:    4, BNC.
  • Computer connection:    USB, Type B socket. Cable supplied.
  • Isolation:    > 250 Vrms independent on each channel.
  • Power:    500 mA @ 5 V DC. Mains adaptor supplied.
  • Dimensions:    170 x 130 x 35 mm.
  • Weight:    ~800 g.
Voltmeter
  • Input impedance:    1013 ohm.
  • Input ranges:    ±2000 mV.
pH & ISE
  • Input impedance:    > 1012 ohm.
  • Input ranges:    ±2000 mV (> pH 0 – 14), ±200 mV (pH 3.6 – 10.4).
  • Calibration:    Single or double point (with Pod-Vu or MF Configurator software).
Conductivity
  • Input ranges:    200, 20, 2, 0.2, 0.02 and 0.002 mS.
  • Calibration:    Single point. Conductivity or TDS.
RTD
  • Probe type:    1000 ohm platinum RTD.
  • Probe error:    ±(0.10 + n/600) °C at n°C with ET021 RTD probe.
  • Temperature range:    –25 to +125°C.
Thermistor
  • Probe type:    30 kohm thermistor.
Dissolved Oxygen dO2
  • Input ranges:    ±2 mA, ±200, ±20, ±2 µA, ±200, ±20 nA.
  • Polarization:    0, then –500 to –1000 mV in 50 mV steps (with Pod-Vu or MF Configurator software).
  • Zero offset range:    ±20 µA.
  • Typical RMS noise:   1 pA when sampling at 1/s.
  • Calibration:    Single or double point (with Pod-Vu or MF Configurator software) .
Biosensor
  • Input ranges:    ±2 mA, ±200, ±20, ±2 µA, ±200, ±20 nA.
  • Polarization range:    ±2000 mV in millivolt steps.
  • Zero offset range:    ±20 µA.
  • Typical RMS noise:    1 pA when sampling at 1/s.
  • Calibration:    Single or double point (with Pod-Vu or MF Configurator software).


 

© copyright 2002 - 2024   eDAQ - data recording made simple